Air Sampling

From MC Chem Wiki
Revision as of 00:36, 18 November 2020 by Jwelker (talk | contribs)
Jump to navigation Jump to search

X-Sorber

Xsorber2.jpg

The X-Sorber is a lightweight sampling device that can be returned to its host instrument for sample desorption, data download, conditioning, and charging. When docked at the host instrument, settings and pre-developed methods can be loaded onto the X-Sorber prior to deployment in the field.

Programming

  • Select File/New/X-Sorber

Setup

  • Parameters on this tab are global properties
  • Default ready temp is 40 degrees C
  • Set point is the threshold above which a sample method will not run
  • Faster flow rate may capture a target of lower concentration, but it may cause analyte breakthrough
  • Range from 150 mL/min to 675 mL/min

Method

  • Click Add in Method tab
  • Enter a method name
  • Select a start type and a sample duration

Start Types

  • Immediate: run starts immediately and is stopped after a programmed duration
  • At time: starts a sample run at a specified future time
  • Delay: starts a sample run at a future time by counting down the time specified by the delay
  • Manual: starts a sample that is physically terminated by the user

Sample Durations

  • Volume: samples a specific volume of air calculated in milliliters
  • Time: samples for a designated time period

Conditioning

System Overview

Features

  • Safe air sampling using dual selectable sorbent tubes
  • Storage of sample information including GPS coordinates, time and date, sample duration and volume, user ID, and tube ID
  • Thermal desorption and sample analysis capabilities
  • Sorbent tube conditioning at the air dock

Sampling and Desorption

  • activation of the sample pump pulls air through the installed sorbent tubes upon which the sample is collected
  • the sorben tube is loaded with the sample's constituent compounds
  • upon docking the X-Sorber to the host instrument, the inlet and appropriate sorbent tube are heated. Carrier gas is then routed in the opposite direction through the flow path, and the desorbed sample components are carried into the GCMS.
  • analytes are adsorbed to the sorbent tube until they cross the breakthrough volume's threshold, causing them to break through the downstream end of the tube.
  • sample collection below the breakthrough volume's threshold allows for proper calculation of the analyte's concentration

Air Dock System

Two dock ports

  • Universal Sampling Port: the main docking port
  • Conditioning Port: can be used for programming and conditioning, but will not transfer samples into the GCMS

Programming at the GCMS

  • sample methods must be programmed at the GCMS using Griffin System Software and must be downloaded on the X-Sorber prior to field use

Conditioning

  • The sorbent tube is heated and a controlled flow of carrier gas is used to desorb any remaining chemical residue

Tube Constraints

  • a maximum of 20 sorbent tube IDs can be recorded on the X-Sorber
  • 10 pre-programmed methods are available
  • each sorbent tube is allowed 10 sample runs before becoming "maxed"
  • tubes should be conditioned and desorbed as close in time to deployment as possible
  • tube lifespans will vary according to sorbent bed material and use

Programming the X-Sorber

  • open griffin system software
  • go to file/new/X-Sorber
  • 40 degrees celsius is the threshold above which a sample will not run
  • flow rate: fasater flow rate may capture a target chemical of lower concentration, but may cause analyte breakthrough. 150-675 mL/min
  • single tube:325-675
  • multi tube:150-350

Inlet Cap

  • inlet cap should be placed over the sampling inlet to cover the pneumatic and electrical ports